MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTXSFT0206
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the carotid terminus using penumbra smart coils (smart coil), a penumbra smart coil detachment handle (handle), a non-penumbra microcatheter, and a non-penumbra embolization device.During the procedure, after successfully implanting a non-penumbra coil in the aneurysm, the physician advanced the smart coil into the aneurysm using the microcatheter and attempted to detach it using the handle; however, the smart coil failed to detach.The physician then attempted to manually detach the smart coil, but the coil still failed to detach.The physician decided to remove the smart coil and microcatheter altogether.Upon retraction, the smart coil unintentionally detached in the parent vessel.Therefore, the physician removed the detached smart coil using a snare device.The procedure was completed at this point.There is no reported adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned smart coil confirmed that the embolization coil was detached.Evaluation revealed that the pusher assembly was kinked and fractured at its proximal end, and the pull wire was fully retracted.This damage likely occurred due to the manual detachment attempts during the procedure, and likely contributed to the reported embolization coil detachment upon retraction.Due to the returned condition, the root cause of the reported complaint could not be determined.Further evaluation of the device revealed that the pusher assembly had additional damage throughout its length.This damage was likely incidental to the complaint and may have occurred during packaging for the device return.Further evaluation also revealed offset coil winds on the embolization coil.This damage likely occurred due to snaring the coil to remove from the patient vessel.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.Section h.Box 6.Conclusions code 1: 4316 - the investigation findings do not lead to a clear conclusion about the root cause of the reported initial detachment issue on the smart coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key13793462
Report Number3005168196-2022-00126
Device Sequence Number3905599
Product Code HCG
UDI-Device Identifier00814548016030
UDI-Public00814548016030
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/22/2026
Device Model Number400SMTXSFT0206
Device Catalogue Number400SMTXSFT0206
Device Lot NumberF109358
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Date Report Sent to FDA01/01/2005
-
-