• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORCYM S.R.L. CARBOMEDICS TOP HAT MECHANICAL HEART VALVE; MECHANICAL HEART VALVE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORCYM S.R.L. CARBOMEDICS TOP HAT MECHANICAL HEART VALVE; MECHANICAL HEART VALVE PROSTHESIS Back to Search Results
Model Number CPHV
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Codes Thrombosis/Thrombus (4440); Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2021, a carbomedics top hat mechanical heart valve s5-023 was implanted.On (b)(6) 2022, the fluoroscopy was performed due to the valve thrombus.A slight valve insufficiency was observed in one of leaflets, but the patient's symptoms were mild and treated with anticoagulant therapy.
 
Additional Manufacturer Narrative
A complete manufacturing and material records review for the device has been performed.The results confirmed that the device satisfied all material, visual and performance standards required at the time of manufacture and release.Since the device remains implanted (no notification of explant received to date), no further investigation is possible at this time.The manufacturer attempted to follow up to retrieve more information, but the manufacturer was informed that no additional information will be provided on the event.Based on the information available, it is not possible to draw a definitive conclusion to the reported event.However, from the document review performed, no manufacturing deficiencies were identified.As reported, the patient's symptoms were mild and no device explant was reported as a result of the event.As such, there is no evidence to suspect a serious adverse event for the patient as a result of the event reported.Given that the patient was treated with anticoagulation therapy, it is possible that the event was caused by an incorrect management of the postoperative anticoagulation therapy.It should be noted that valve thrombosis is listed as a possible adverse event in the carbomedics prosthetic heart valve ifu.The event is, therefore, a known inherent risk of the device.Should any additional information be received in the future, a follow up report will be provided.Device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBOMEDICS TOP HAT MECHANICAL HEART VALVE
Common Device Name
MECHANICAL HEART VALVE PROSTHESIS
Manufacturer (Section D)
CORCYM S.R.L.
strada crescentino
saluggia, vercelli 13040
IT  13040
Manufacturer (Section G)
CORCYM S.R.L.
strada crescentino
saluggia, vercelli 13040
IT   13040
Manufacturer Contact
francesca crovato
5005 north fraser way
burnaby, bc V5J 5-M1
CA   V5J 5M1
MDR Report Key13793676
Report Number3005687633-2022-00111
Device Sequence Number1765434
Product Code LWQ
UDI-Device Identifier08022057012913
UDI-Public(01)08022057012913(240)S5-023(17)240311
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P900060
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/11/2024
Device Model NumberCPHV
Device Catalogue NumberS5-023
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2022
Initial Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age54 YR
Patient SexMale
-
-