MAUDE Adverse Event Report: ENCORE MEDICAL L.P FOUNDATION HIP; STEM, HIP,FH 470 SERIES SIZE12 PC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P FOUNDATION HIP; STEM, HIP,FH 470 SERIES SIZE12 PC Back to Search Results
Model Number 470-01-120
Health Effect - Clinical Code Fall (1848)
Date of Event 02/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery: due to patient fell and fractured.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as fracture after a fall.The previous surgery and the surgery detailed in this event occurred 24 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr# (b)(4) associated with the main part #470-01-120, stem, hip, fh 470 series size12 pc which document that out of 10 parts lot, 6 parts were rejected due to thread gauge does not go to min depth.Later, the rejected parts were returned to vendor.There was an another ncmr# (b)(4) associated with the same part which document that out of 10 parts lot, 6 parts were rejected and scrapped due to thread gauge does not go to min depth.Later the rejected parts were reworked and accepted after proper justification.All other items in the respective lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to fracture after a fall.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Agent has clearly mentioned that "patient fell" and due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions, incorrect implant selection, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION HIP
Common Device Name
STEM, HIP,FH 470 SERIES SIZE12 PC
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key13793689
Report Number1644408-2022-00331
Device Sequence Number1765439
Product Code LZO
UDI-Device Identifier00888912086264
UDI-Public(01)00888912086264
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K991226
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/23/2025
Device Model Number470-01-120
Device Catalogue Number470-01-120
Device Lot Number564C1921
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2022
Initial Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
412-02-052 LOT 000C1115; 497-28-000 LOT 629C1676
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
-
-