DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/18/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that before use the cardiosave intra-aortic balloon pump (iabp) was having a battery charging issue, the unit was running on battery mode even when plugged in.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) evaluated the iabp, but was unable to reproduce the reported "battery charging" issue.The fse then performed preventative maintenance (pm) including a full calibration, functional testing and safety checks to factory specifications.All testing passed and the iabp unit was returned to the customer and cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.
|
|