Model Number RT041
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A distributor reported on behalf of a healthcare facility in (b)(6) reported that the seal of ten rt041 full face masks disconnected from the mask frame.There was no reported patient involvement.
 
Additional Manufacturer Narrative
(b)(4).The complaint rt041 full face masks are currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Additional Manufacturer Narrative
(b)(4).Method: the complaint rt041 non-vented hospital full face masks were returned to fph in new zealand and were visually inspected.Results: visual inspection of the returned devices revealed that the seal were partly separated from the mask base.However, a continuous bead of glue were found around the masks seals.Further inspection revealed that the part of the seals that were not separated from the bases indicates that the seals had been properly inserted in the mask bases.Conclusion: we were unable to determine the root cause of the observed separation.Visual inspection indicated that the masks were assembled and glued properly.This suggests that the separation occurred post production.The rt041 full face mask features a mask base, seal, and headgear.The mask seal is inserted into the mask base and is secured with glue.Each assembled mask seal is inspected to ensure it is correctly inserted and there are no unacceptable gaps between the seal and the base.Samples are taken and the mask seals are pulled apart after 24 hours to ensure they meet or exceed the required detachment force before the assembled product can be released.Our user instructions illustrate in pictorial format the correct set-up and proper use of the rt041 non-vented hospital full face mask.It also states the following: - "operating pressure range: 5 - 25 cmh2o".- "this mask may only be used in a hospital or clinical setting where the patient is adequately monitored by trained medical staff.Failure to monitor the patient may result in loss of therapy, serious injury or death.".- "in the case of therapy device failure, the use of this mask requires the same level of attention and assistance as in the use of a tracheal tube.".
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in colorado reported that the seal of ten rt041 full face masks disconnected from the mask frame.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NON-VENTED HOSPITAL FULL FACE MASK
Common Device Name
CBK
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key13794887
Report Number9611451-2022-00199
Device Sequence Number4722835
Product Code CBK
UDI-Device Identifier09420012429858
UDI-Public(01)09420012429858(10)2101709894(11)210707
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K083122
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRT041
Device Catalogue NumberRT041
Device Lot Number2101709894
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer05/19/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-