Model Number FX415T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigations: visual inspection: no significant deformations or damage of the valves were detected during the visual inspection.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: the valves are tested in both the horizontal as well as the vertical positions.The results show that the valves operate within the accepted tolerance in both positions.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Klick force and brake function test: the brake functionality test has shown that the brake function is fully operational and the klick force is within the given tolerances.Internal inspection: after dismantling of the valves, clearly deposits were found in both valves.Results: we would like to point out in advance that after the revision operation there was already done the permeability and adjustment test.At the time of the investigation, it was not clear to us how the mentioned functional impairment occurred.No functional deviations were detected.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of non-visible deposits/proteins might compromise the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke (b)(4).No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav 2.0 shuntsystem (part # fx415t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt system was believed to be operated in overdrainage and was difficult to adjust.The patient underwent a revision procedure on an unknown date.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6).Height: 190 centimeters (cm).Weight: (b)(6).Gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 30
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13795092
Report Number3004721439-2022-00047
Device Sequence Number17401564
Product Code JXG
UDI-Device Identifier04041906136574
UDI-Public4041906136574
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/04/2022
Device Model NumberFX415T
Device Catalogue NumberFX415T
Device Lot Number20035014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2022
Initial Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/05/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age18 YR
Patient SexMale
Patient Weight85 KG
-
-