AMO PUERTO RICO MFG. INC. TECNIS CL IOL; INTRAOCULAR LENS Back to Search Results
Model Number Z9002
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the intraocular lens (iol) was explanted from patient's left eye as the lens decentered.There was no patient injury, however, a vitrectomy and iridotomy were performed.Another competitor lens was used as the replacement.The patient's post-operative outcome was not available.No further information was provided.
 
Additional Manufacturer Narrative
Pt info: unknown/asked but unavailable.Initial reporter: first/given name: unknown, information not provided.Device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: mar 23, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received inside a specimen cup submerged in an unknown solution.Visual inspection under magnification revealed that the lens was returned cut in half.No other defects were observed.Based on the return condition of the lens (cut in half) no further product evaluation could be performed on the lens.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13795097
Report Number3012236936-2022-00634
Device Sequence Number5032748
Product Code HQL
UDI-Device Identifier05050474530706
UDI-Public(01)05050474530706(17)170201
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2017
Device Model NumberZ9002
Device Catalogue NumberZ900200205
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer04/25/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-