| Model Number |
TMICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Sex: unk, weight: unk, ethnicity: unk, race: unk, date of event: unk, explant date: unk, work order search: another similar complaint was reported for units within the same lot.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens, -05.00/+1.5/090 (sphere/cylinder/axis), in the patients left eye (os) on (b)(6) 2022.The lens was explanted due to an over vault.The lens was replaced with a shorter lens.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: the lens was explanted on (b)(6) 2022 due to excessive vaulting.The lens was replaced with a shorter lens and the problem was resolved.The cause of the event was due to the device.Claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|