MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HD AUTOCLAVABLE CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HD AUTOCLAVABLE CAMERA HEAD Back to Search Results
Model Number OTV-S7PROH-HD-12Q
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2021
Type of Reportable Event Malfunction
Event or Problem Description
Olympus customer care was informed that the customer high definition autoclavable camera head will be returned for a reported ¿black screen with no image¿ found during preparation for use.No patient injury or harm was reported to olympus.
 
Additional Manufacturer Narrative
The device was returned to the service center for evaluation.The customer¿s reported issue was confirmed as no image was displayed due to a defective charged coupled device (ccd) unit.Additionally, the cable unit has deformation/kinks.The cause of the defective ccd is unknown.The investigation is ongoing; therefore, the root cause of the reported issue cannot be determined at this time.However, if additional information becomes available a follow up medical device report will be supplemented accordingly.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the faulty charge coupled device could not be determined.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: "do not coil the camera cable with a diameter of less than 20 cm.The camera cable may be damaged.Never excessively pull the camera cable, but straighten it gradually when it is coiled.The camera cable could be damaged.Never excessively bend, pull, twist, coil, squeeze, or crush the camera cable.The camera cable could be damaged.Do not use excessive force when wiping the external surfaces of the camera cable.The camera cable could be damaged.Never attempt to lift the entire assembly by the camera cable while the camera head is attached to the endoscope.The camera cable could be damaged.Never use a clamp or forceps to attach the camera cable to another object.The camera cable could be damaged." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD AUTOCLAVABLE CAMERA HEAD
Common Device Name
CAMERA HEAD
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13796003
Report Number8010047-2022-04582
Device Sequence Number5063089
Product Code GCJ
UDI-Device Identifier04953170228865
UDI-Public04953170228865
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K955404
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOTV-S7PROH-HD-12Q
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2021
Supplement Date Received by Manufacturer03/24/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/20/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-