PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker evaluated the customer's device and verified the reported issue.Stryker replaced the device's main keypad to resolve the reported issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing and the device returned to the customer for use.Stryker further evaluated the removed main keypad and contamination between dome switch and carbon trace of buttons (display, pause, shock, right side charge and energy up) was determined to be the cause of the reported issue.Contaminant contains phosphorus, sodium, sulfur and chlorine which are common clean product ingredients.
 
Event or Problem Description
The customer contacted stryker to report that their device was not delivering defibrillation energy.Customer further advised the device's shock button was not functioning.There was no report of patient use associated to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key13796041
Report Number0003015876-2022-00384
Device Sequence Number20091814
Product Code MKJ
UDI-Device Identifier00883873872079
UDI-Public00883873872079
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2022
Initial Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-