| Model Number |
C2A12 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product is being returned to applied medical for evaluation but lot number is provided.A follow-up report will be provided upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: vnotes workshop.Event description: cer from vnotes workshop.The following information was provided on 18 february via email by [name]: i need to update you on a product issue.I ran a vnotes workshop today in adelaide with [name] and [name] as faculty.[name] was teaching a couple of consultants with our gelpoint vpath when a suture (see first picture below for details) was introduced into one of the 10mm sleeves, and part of the device popped out and into the simsei.[name] is concerned as the plastic was pushed out from the sleeves.Patient status: na.
|
| |
|
Event or Problem Description
|
|
Procedure performed: vnotes workshop: event description: cer from vnotes workshop the following information was provided on 18 february via email by [name]: i need to update you on a product issue.I ran a vnotes workshop today in adelaide with [name] and [name] as faculty.[name] was teaching a couple of consultants with our gelpoint vpath when a suture (see first picture below for details) was introduced into one of the 10mm sleeves, and part of the device popped out and into the simsei.[name] is concerned as the plastic was pushed out from the sleeves.Patient status:.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was not returned to applied medical for evaluation.However, a photo of the shield, an internal plastic component of the sleeve, was provided, confirming the complainant¿s experience of the shield dislodging from the sleeve.Based on the description of the event, it is likely that the reported event was caused by non-axial insertion or removal of asymmetrical instrumentation (such as the suture needle) through the sleeve.Applied medical¿s instructions for use (ifu) states that, "extra care should be taken when inserting and withdrawing angular and asymmetrical instruments, such as 'j' hooks and clip appliers." in addition, the instructions for use (ifu) also states that, "all instruments should be centered axially when inserted through the seal.".
|
| |
|