• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227)
Date of Event 02/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the preloaded intraocular lens (iol) was explanted from patient's left eye due to glare and halos.A competitor's lens was used as the replacement.There was no patient injury and no other intervention was required.No further information was provided.
 
Additional Manufacturer Narrative
Weight and ethnicity: unknown, information not provided.Date of event: unknown, not provided, but the best estimate date is between aug 4, 2021 and feb 17, 2022.The device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: per form that came in with the product return, date of event was feb.17, 2022.Field below updated to reflect this: section b3: date of event: feb.17, 2022.Section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: mar 23, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the lens was received cut into three pieces (w/ one piece missing).The lens was cleaned and, no additional issues were observed.Based on the return condition of the lens no further product evaluation could be performed on the lens.No handpiece was received as part of this return.Based on the complaint investigation results, the complaint issue was not confirmed.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13797494
Report Number3012236936-2022-00637
Device Sequence Number17169930
Product Code HQL
UDI-Device Identifier05050474750579
UDI-Public(01)05050474750579(17)240306
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/06/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU150
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date05/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-