ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF RESTOR TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SND1T5
Device Problem Failure to Align (2522)
Patient Problems Intraocular Pressure Increased (1937); Blurred Vision (2137)
Event Date 02/16/2022
Event Type  Injury  
Event Description
A healthcare professional reported that following an intraocular lens (iol) implant procedure, the lens was rotated out of axis and repositioning has been planned.She had an intraocular pressure (iop) of 29 and was given lumigan to use twice a day for 1 week to reduce the pressure.At visit 2, the patient reported decreased vision.After dilation, the patient was examined by the doctor.He determined the iol to be positioned at 75° axis.Additional information was requested.
 
Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: 2022-10659.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF RESTOR TORIC SINGLEPIECE IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13798592
MDR Text Key287351801
Report Number1119421-2022-00480
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberSND1T5
Device Catalogue NumberSND1T5.180
Device Lot Number15059938
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient SexFemale
-
-