| Model Number |
7540220 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a distributor and healthcare provider (hcp) via a medtronic representative regarding a patient implanted with screws in a posterior lumbar interbody graft fusion for a lumbar disc herniation.It was reported that when the screws were implanted, they were screwed into the wrong thread causing the screws to slip.The screws had been in contact with the patient during the patient's surgical implantation process.When it was found that the screws slipped, they were removed.There was no additional surgery or treatment/ revision surgery/ delay in overall procedure time/ hospitalization necessary as a result of this event.The event was associated with a patient. there were no patient symptoms or complications as a result of this event.There was no treatment or additional surgery performed as a result of this event.The dates of implant and explant was (b)(6) 2022.No further complications were reported/ anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis part# 7540220 ; lot# h5661498- visual and optical inspection confirmed the screw was returned with slight thread damage.The thread crest damage appears to be from the misalignment of the set screw when attaching and threading in the bone screw.The female torx has also been deformed.This damage may have happened during the removal process.H6: updated eval.Code method and eval.Code result post analysis medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|