| Model Number |
C111 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer changed the sample probe and increased the extra washes and wash volume.After service, the customer ran five samples with no issues.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable high ggt-2 gamma-glutamyltransferase ver.2 results from the cobas c111 analyzer.Sample 1 initial result was 138.4 ku/l and the repeat result was 9.1 ku/l.From a cobas c501 analyzer, the result was 11 ku/l.On (b)(6) 2022, sample 2 initial result was 87.5 ku/l and the repeat result was 26.5 ku/l.From a cobas c501 analyzer, the result was 23 ku/l.The questionable results were not reported outside of the laboratory.The reagent lot number and expiration date were requested but were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|