• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5532-G-411
Medical Device Problem Codes Material Discolored (1170); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cyst(s) (1800); Pain (1994)
Date of Event 02/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's left knee was revised due to pain.Surgeon performed cyst removal and exchanged a 4x11 ps insert for a 4x12 ps insert.Upon explantation, the insert was seen to have some form of discoloration/ crystallization under the surface of the insert.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that the patient's left knee was revised due to pain.Surgeon performed cyst removal and exchanged a 4x11 ps insert for a 4x12 ps insert.Upon explantation, the insert was seen to have some form of discoloration/ crystallization under the surface of the insert.
 
Additional Manufacturer Narrative
Reported event: an event regarding pain involving a triathlon insert was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection: visual inspection was performed as part of the material analysis: image shows the articulating surface of the triathlon ps x3 tibial insert.Slight yellow discolouration consistent with oxidation, possibly a result of synovial fluid absorption, was observed on the insert.Explantation damage was observed on the distal surface of the insert.Explantation damage was observed on the anterior surface of the insert.Material analysis: a material analysis was performed and concluded the following: "review of triathlon ps x3 tibial insert, catalogue # 5532-g-411, lot code tr6kke confirmed the presence of discolouration.Characterisation using stereo microscopy confirmed discolouration consistent with absorption of synovial fluid.No manufacturing related defects were observed on the device surfaces examined." clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to pain.It was further reported that the device appeared discolored upon explantation.A material analysis was performed on the returned device and concluded the following: "review of triathlon ps x3 tibial insert, catalogue # 5532-g-411, lot code tr6kke confirmed the presence of discolouration.Characterisation using stereo microscopy confirmed discolouration consistent with absorption of synovial fluid.No manufacturing related defects were observed on the device surfaces examined." the exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick, NJ NA
EI   NA
2018315000
MDR Report Key13800899
Report Number0002249697-2022-00409
Device Sequence Number4030065
Product Code MBH
UDI-Device Identifier07613327050714
UDI-Public07613327050714
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model Number5532-G-411
Device Catalogue Number5532-G-411
Device Lot NumberTR6KKE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer02/24/2023
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexFemale
-
-