• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 1076538
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Event or Problem Description
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.The complaint includes ac/dc connector damaged and blower is noisy, tobacco degradation, ac/dc connector, blower assembly, valve assembly, blower cap, blower housing, bottom enclosure, internal air cover, internal foam, blower box mount, right panel assembly, pollen filter need to be replaced.There was no report of serious or permanent patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
The manufacturer previously reported that hospitalization - initial or prolonged, disability or permanent damage occured in section b2.After review, it was determined that all these were incorrectly selected in section b2.Nothing such has occured to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13801196
Report Number2518422-2022-10519
Device Sequence Number4005983
Product Code BZD
UDI-Device Identifier00606959060302
UDI-Public00606959060302
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1076538
Device Catalogue Number1076538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2018
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/04/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Disability;
Patient SexUnknown
-
-