| Model Number |
03P6822 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a (b)(6) magnesium result generated on the architect c4000 analyzer on a patient.Results provided: (b)(6) 2022 (b)(6).Normal range: 1.6-2.6 mg/dl.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Ticket searches and performance testing could not be performed as the likely cause lot number is unknown.There are no trends for the product related to patient results.No customer returns were available for evaluation.A review of overall customer field data was performed to assess the performance of the architect magnesium assay and is within the established baselines confirming no systemic issues in the field.A review of the manufacturing documentation did not identify any issues associated with the customers observation.A review of labeling concluded that the issue is sufficiently addressed.No product deficiency was identified.
|
| |
|