|
Investigation ¿ evaluation: (b)(6) (united states) contacted cook on (b)(6) 2022.Stating that, at the end of a drain exchange procedure when wire guide was removed, the wire guide was unraveled.The wire guide is a thsf-35-125-thg from an unknown lot.The wire guide was not withdrawn through a needle at any point but was withdrawn out of the catheter.The wire guide insertion tool was not used, however there was no difficulty inserting the wire guide into the patient.It is not known if the product was torqued.There were no adverse effects for the patient due to the wire not being used.The patient did not experience any additional procedures due to this occurrence.The date of event for this occurrence was 28feb2022.Reviews of the complaint history, device history record (dhr), quality control, manufacturing instructions (mi), and instructions for use (ifu), as well as a visual inspection of the returned device, were conducted during the investigation.The product was returned to cook for evaluation.One used wire guide was received.The proximal weld connection was separated from the inner mandrill.As a result, mandrill protrusion occurred.The mandrill was sticking out of the coil at 7.7cm.The proximal weld was present and attached to coil.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The customer was not able to clarify the lot number for this complaint; however, two possible lot numbers (14385939 and 14411855) were provided.A device history record for these two lots was performed.Lot# 14385939 had one relevant non-conformance where 3 devices were scrapped due to incorrect solder length.Lot# 14411855 had three relevant non-conformances.The first was for 1 device being scrapped due to weld splash in unspecified area.The second was for one device scrapped due to a broken solder.The third was for two devices scrapped due to stretched coil.A database search did not identify any other events associated with the two possible lot numbers.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in field.Cook also reviewed product labeling.This product is supplied with an instructions for use (ifu) pamphlet, t_fcwg_rev1.In the warnings section it states: this product is a delicate instrument.Avoid forceful angulation.In the precautions section it states: when you use the wire guide with another device, consider the end-hole size and the length of the device in order to ensure a proper fit between the wire guide and the device.Use only the flexible end as the initial inserted end.Inserting from the rigid end may cause damage to tissue or the device.Inspect the wire guide for kinks or damage prior to use.In the instructions for use section, it states: 3.) if needed, insert a wire guide insertion tool (provided) through the valve assembly or hub of the guiding sheath or other interventional device.Insert the tip of the wire guide through the insertion tool.In the how supplied section it states: upon removal from package, inspect the product to ensure no damage has occurred.Based on the available information, inspection of the returned device, and the results of the investigation, a definitive root cause was unable to be established.There is slight coil elongation, less than 1mm, at the bottom of the first anchor solder, closest to the distal weld.This failure could be due to component failure or manipulation of the wire during removal.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|