It was reported that the blue rhino dilator included in the blue rhino g2-multi percutaneous tracheostomy introducer set was unable to be advanced during a percutaneous tracheostomy procedure on a male patient.As the physician advanced the blue rhino dilator, it was "unable to pass" while creating the stoma.It was noted that the distal end of the dilator was bent, making dilation impossible.The tracheotomy was then completed with a tracheostomy kit from another manufacturer.The patient did not experience any adverse effects or require additional procedures due to this occurrence.Additional information regarding event details has been requested, but is currently unavailable.
|
|
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Investigation ¿ evaluation hosp.(b)(6) (spain) informed cook of an incident involving a c-ptis-100-hc-g-eu-flex8.5 (blue rhino g2-multi percutaneous tracheostomy introducer set) from lot 13848230.It was reported, that while the user was performing a percutaneous tracheostomy with a blue rhino set, the user was unable to pass the blue rhino dilator to perform the stoma.The patient was described as elderly with possible tracheal calcification.The procedure was then completed using a different product.A review of the complaint history, device history record, and quality control were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record.Cook concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the device history record for lot 13848230 to check for related nonconformances and additional complaints.Lot 13848230 did not have related nonconformances.Pti-38.0-hc-g dilator component from lot ic13687665 did not have nonconformances.There are no additional complaints on lot 13848230.There is no evidence of nonconforming material in house or in the field.The information provided upon review of the device master record and device history record, does not indicate the device was manufactured out of specification.Cook also reviewed product labeling.The ifu packaged with the device contains the following in relation to the reported failure mode: "instructions for use: patient preparation 6.Generously lubricate the surface of the appropriately sized loading dilator and load the tracheostomy tube onto the dilator.Ensure that the tracheostomy tube's tip fits snugly on the dilator.Ensure that the balloon is completely deflated.Thoroughly lubricate tracheostomy tube assembly.Tracheostomy procedure: 2.After introducing local anesthesia, make a 1.5-2.0 cm skin incision (vertical or horizontal) at the chosen insertion site.3.If desired, use a curved mosquito clamp to gently dissect vertically and transversely down to the anterior tracheal wall.With a finger tip, dissect the front of the trachea, in the midline, free of any tissues and identify the cricoid cartilage.Displace the isthmus of the thyroid downward, if present.How supplied upon removal from package, inspect the product to ensure no damage has occurred." based on the information provided and the results of the investigation, cook has concluded that component failure unrelated to manufacturing or design deficiencies contributed to this incident.It should also be noted the customer stated the patient was elderly and had possible tracheal calcification.This may have been a factor in this failure, but this cannot be definitively confirmed.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|