| Catalog Number |
59038UAC |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device will not be returning for evaluation.A lot number was reported, and a review of the manufacturing history record is in progress.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a uterine fibroid embolization procedure, the tip of the diagnostic 5f catheter detached within the patient.The physician had acquired retrograde femoral arterial access, and during catheter manipulations, over a guidewire, the catheter tip detached within the patient.The physician attempted to retrieve the foreign body with a vascular snare device but was unsuccessful.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|