Catalog Number 241.901
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 02/04/2022
Event Type  Injury  
Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, after a pose of a philos plate, a recovery surgery was performed, and the surgeon observed the absence of locking of the proximal screws intra-operatively.The plate was ablated and was replaced during a same recovery surgery.This report is for one (1) humerus plate.This is report 1 of 2 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the philos 3.5 3ho sst.A dimensional inspection for the philos 3.5 3ho sst was unable to be performed due to complex geometry of the device.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the philos 3.5 3ho sst was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H6: a device history record (dhr) review was conducted: part: 241.901, lot: 308p188, manufacturing site: raron, release to warehouse date: 05.Aug.2021.A manufacturing record evaluation was performed for the finished device 241.901, lot: 308p188 and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5MM LCP PROXIMAL HUMERUS PLATE-STANDARD 3H SHAFT/90MM
Type of Device
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, SINGLE COMPONENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK RARON (CH)
kanalstrasse west 30
raron 3942
SZ   3942
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13803078
MDR Text Key291405873
Report Number8030965-2022-01705
Device Sequence Number1
Product Code KTW
UDI-Device Identifier07611819115071
UDI-Public(01)07611819115071
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K011815
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number241.901
Device Lot Number308P188
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - SCREWS: PHILOS
Patient Outcome(s) Required Intervention;
-
-