• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC SPEEDSHIFT COMP IMPL KIT 15X20X20 OFST 8; STAPLE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC SPEEDSHIFT COMP IMPL KIT 15X20X20 OFST 8; STAPLE, FIXATION, BONE Back to Search Results
Model Number SE-1520-08
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that, the speed shift compression implant kit staples disengaged from the handles.No patient and procedure involvement.This report is for (1) speedshift comp impl kit 15x20x20 ofst 8.This is report 2 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: part # se-1520-08.Synthes lot # mse210173.Supplier lot # mse210173.Release to warehouse date: september 29, 2021.Expiration date: june 1, 2026.Supplier: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the photo.Visual analysis of the photo revealed that the speedshift comp impl kit 15x20x20 ofst 8 had the staple unattached inside the sealed package, this condition is most likely due to breakage of the clamps, staple and fragment are visible in the evidence provided.The observed condition of the speedshift comp impl kit 15x20x20 ofst 8 was consistent with a design related issue.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the speedshift comp impl kit 15x20x20 ofst 8.The potential impact of the design to the complaint condition is addressed under a capa.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that speedshift comp impl kit 15x20x20 ofst 8 was broken a part of the insertion stick body implant.The device condition can be observed to be broken and deployed inside its original sealed packaging.The dimensional inspection was not performed as this is a known issue addressed under a capa.The observed condition of the speedshift comp impl kit 15x20x20 ofst 8 was consistent with a design related issue.A functional test was not performed as the device was found to be broken inside the original packaging.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the spedshift comp impl kit 15x20x20 ofst 8 was observed to broken and deployed inside the original packaging.The potential impact of the design to the complaint condition is addressed under a capa.No further corrective action is deemed necessary at this time.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4: lot part #: se-1520-08 synthes lot #: mse210129 supplier lot #: mse210129 release to warehouse date: 10 aug 2021 expiration date: 01 mar 2026 supplier: (b)(4), no ncr's generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDSHIFT COMP IMPL KIT 15X20X20 OFST 8
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES BME : SAN ANTONIO
14785 omicron drive
#205
san antonio TX 78245
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13803356
Report Number2939274-2022-00923
Device Sequence Number14460553
Product Code JDR
UDI-Device Identifier00810633020432
UDI-Public(01)00810633020432
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K142292
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2026
Device Model NumberSE-1520-08
Device Catalogue NumberSE-1520-08
Device Lot NumberMSE210129
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer03/30/2022
04/04/2022
05/11/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date03/30/2022
04/13/2022
06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SPEEDSHIFT COMP IMPL KIT 15X20X20 OFST 8
Patient SexUnknown
-
-