CARDINAL HEALTH SALEM SUMP W/ ARV 16FR. 48IN; TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION Back to Search Results
Model Number 8888266130
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the device has been discarded.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
Customer reports: after inserting the tube into the patient the nurse noticed that it was not placed properly.When they tried to remove the tube they were unable to because it was stuck and inflicting pain on the patient.They cut the tube and removed some through the mouth and the rest through the nose.The tube somehow knotted itself while it was being inserted.
 
Additional Manufacturer Narrative
The device history record review could not be performed since no lot number was provided from customer complaint report, and no photo sample/actual sample with an identifiable lot number was received.The complaint report indicates that no sample is available in connection with this complaint report however, two photographs were submitted for review and investigation.After a visual inspection of the photos, it appears that there is a knot in the catheter.A gemba walk through was conducted in the manufacturing area with the multifunctional team (quality, manufacturing, engineering); it was determined that the potential root cause could not have happened in the manufacturing floor as there is no step in the manufacturing process in which the product is handled to the point of knotting the device.A visual inspection is performed before the release of the product with acceptable results.Based on the analysis, the observed occurrence of harm improbable which is below the range of the expected occurrence rate therefore corrective actions are not deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SALEM SUMP W/ ARV 16FR. 48IN
Common Device Name
TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION
Manufacturer (Section D)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key13803950
Report Number9612030-2022-03192
Device Sequence Number20091863
Product Code FEG
UDI-Device Identifier10884521004719
UDI-Public10884521004719
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8888266130
Device Catalogue Number8888266130
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-