MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-N180
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus, the evis exera ii gastrointestinal videoscope tested positive for 10 colony forming unit (cfu) of mixed spp.The instrument channel was sampled.The issue was found during a routine culture of the scope.Sampling was taken at reprocessing, before use.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
The olympus scope was sent to an independent laboratory for culture testing.All channels were sampled.There was no detection of micro-organisms presents on the scope and in the channels.The user facility provided additional information regarding the cleaning, the disinfection and the sterilization processes performed onsite for the endoscopes.During pre-cleaning, the customer water was aspirated though the instrument/suction channel with a suction pump until the aspirated liquid became clear, forceps elevator moved to raise and lower 3 times in water during aspiration, and the auxiliary channel flushed with water.There was a less than 60 mins transition period between precleaning to manual cleaning.During manual cleaning, leak testing was performed according to the ifu.The customer used dr weigert mediclean forte detergent with channel-opening maj-1888 brush to brushed and flushed the air/water channel, instrument/suction channel, suction cylinder, instrument channel port and the auxiliary channel.In addition, the forceps elevator was operated (up/down) in the detergent solution.There was no manual disinfects of the scope performed.For automated endoscope reprocessor (aer) treatment, the getinge ew2 reprocessor along with getinge pokayoke detergent and getinge pokayoke a + b disinfectant were used.The scope was stored in a drying cabinet for 30 days with all removable parts detached and olympus provided the processing training.The scope was not sterilized.The customer stated that no automated endoscope reprocessor testing was performed.The customer suspected device was returned for investigation.Upon evaluation of the returned device the following defects were noticed: insertion tube and protector (chemical damage, delamination, surface damage, staining, buckle, sneakiness, peel-off, cuts, flexibility), and foreign material at the distal end unit due to insufficient cleaning of the scope.The faulty parts were replaced and device return to the user facility.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "warning: an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13803963
Report Number8010047-2022-04607
Device Sequence Number20092006
Product Code FDS
UDI-Device Identifier04953170202261
UDI-Public04953170202261
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIF-N180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer04/19/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/16/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-