Model Number 3.2-EPR-COL-OP
Health Effect - Clinical Code Perforation (2001)
Date of Event 01/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Target area for er resection: transverse colon, 2cm x 0.5cm scarred tissue bed total time for emergency room resection: 15min.Orientation of scar bed to scope: length of scar bed was east to west (3 > 9 o'clock on a face of a clock) physician noted a 'hole' in the colon tissue.Upon going back to 9 o'clock section to continue resecting, it was noted there was a 'hole' at 7-8 o'clock.Emergency room catheter was immediately removed from scope and 10 clips were used to close the 5-6mm perforation site 10 clips were used to close the perforation.Pt was stable.The antibiotics and ct had been ordered and patient was to be admitted for observation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOROTOR
Common Device Name
ENDOSCOPIC MORCELLATOR
Manufacturer (Section D)
INTERSCOPE, INC
200 commerce drive floor 2
northbridge MA 01534
Manufacturer Contact
jeffery ryan
200 commerce drive floor 2
northbridge, MA 01534
8774207299
MDR Report Key13804529
Report Number3011697028-2022-00001
Device Sequence Number10980519
Product Code PTE
UDI-Device Identifier00854172008131
UDI-Public(01)00854172008131(17)230708(10)210708A
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/08/2023
Device Model Number3.2-EPR-COL-OP
Device Catalogue Number3.2-EPR-COL-OP
Device Lot Number21078A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Initial Report FDA Received Date03/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-