WRIGHTS LANE SYNTHES USA PRODUCTS LLC 12.0MM REAMER HEAD FOR RIA 2 STERILE Back to Search Results
Model Number 03.404.020S
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional device product codes: hrx.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2022, while attempting to pass the reamer head for reamer / irrigator / aspirator (ria) 2 became dislodged after the splines broke.The reamer was removed and the surgeon asked for a smaller one.Fragments of the ria head remained in the femoral canal.There were ten minutes surgical delay.The procedure was successfully completed.This report involves (1) 12.0mm reamer head for ria 2 sterile.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: manufacturing location: supplier ¿ mark two engineering / inspected and packaged by: monument, release to warehouse date: 26-aug-2021, expiration date: 31-jul-2031, part number: 03.404.020s, 12mm reamer head for ria 2-sterile, lot number: 344p090 (sterile).Production order traveler met all inspection acceptance criteria.Packaging label log (pll), was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Component part(s) reviewed: part number: 03.404m020, ria ii reamer head 12.0mm, lot number: 80p5320.Inspection instruction met all inspection acceptance criteria.Inspection sheet, incoming final inspection, met all inspection acceptance criteria.Certificate of compliance was reviewed and determined to be conforming.Certification for heat treat was reviewed and determined to be conforming.Certificate of analysis was reviewed and determined to be conforming.Raw material certification was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria with no issues documented during the inspection or release of the product that would contribute to this complaint condition.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the 12.0mm reamer head for ria 2 sterile was found the device broken.No other defect found.The complaint of fragment inside of patient cannot be confirmed with the evidence available.A dimensional inspection for the 12.0mm reamer head for ria 2 sterile was unable to be performed due to post manufacturing damage.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 12.0mm reamer head for ria 2 sterile would contribute to the complained device issue.Due to the trend with the broken ria 2 reamer heads identified during post market surveillance, the capa was raised to determine the root cause of broken head breakages.Additionally, the product issue escalation was initiated to define any further action related to the breakage.No new malfunctions were observed during this investigation that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12.0MM REAMER HEAD FOR RIA 2 STERILE
Common Device Name
REAMER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13804673
Report Number2939274-2022-00928
Device Sequence Number4029011
Product Code HTO
UDI-Device Identifier10886982273987
UDI-Public(01)10886982273987
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K111437
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.404.020S
Device Catalogue Number03.404.020S
Device Lot Number344P090
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer04/04/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0575-2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - REAMERS
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
Patient Weight58 KG
-
-