Model Number H7493932800350
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a catheter perforation occurred.The patient presented with leg swelling for a venogram with a bilateral iliac stent.As the oc35 peripheral ivus catheter was being loaded on the guidewire, the back end of the wire punctured a hole in the catheter instead of coming out the hub.The device was removed from the wire and the procedure completed successfully with another of the same device.There were no patient complications.
 
Additional Manufacturer Narrative
Pro code (product code) - obj, itx.
 
Additional Manufacturer Narrative
D2b pro code (product code) - obj, it.The device was returned for analysis.Visual inspection revealed a kink in the telescope assembly and a perforation in the guidewire lumen begging 0.70cm from the strain release.Microscopic inspection revealed pitting and degradation of the transducer housing consistent with saline damage which occurs post-use of the device and is not related to the original device failure.
 
Event or Problem Description
It was reported that a catheter perforation occurred.The patient presented with leg swelling for a venogram with a bilateral iliac stent.As the oc35 peripheral ivus catheter was being loaded on the guidewire, the back end of the wire punctured a hole in the catheter instead of coming out the hub.The device was removed from the wire and the procedure completed successfully with another of the same device.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS 35
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13805344
Report Number2134265-2022-03074
Device Sequence Number12053187
Product Code OBJ
UDI-Device Identifier08714729984542
UDI-Public08714729984542
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K200733
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/07/2022
Device Model NumberH7493932800350
Device Catalogue NumberH7493932800350
Device Lot Number0028557742
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer04/12/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient RaceWhite
-
-