• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DJO LLC SENIOR SOLUTIONS 4CH STIM PKG US STD; STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DJO LLC SENIOR SOLUTIONS 4CH STIM PKG US STD; STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT Back to Search Results
Model Number 2866
Health Effect - Clinical Code Superficial (First Degree) Burn (2685)
Date of Event 02/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that the stim unit allegedly burned the patient.Djo is awaiting the return of this device.Additional reporting on this event will be provided as a supplemental report to this document if and when the device is returned for evaluation.
 
Event or Problem Description
It was reported that the stim unit allegedly burned the patient.
 
Additional Manufacturer Narrative
After performing functional and electrical testing, the technician was unable to duplicate the reported complaint.The product met manufacturer specifications.Possible root causes include moisture or misapplication of the electrodes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENIOR SOLUTIONS 4CH STIM PKG US STD
Common Device Name
STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT
Manufacturer (Section D)
DJO LLC
5919 sea otter place
ste 200
carlsbad CA 92010
Manufacturer (Section G)
DJ ORTHOPEDICS DE MEXICO, S.A. DE C.V.
carretera libre tijuana tecate
20230 parque industrial el flo
tijuana, b.c. 22245
MX   22245
Manufacturer Contact
jim pomeroy
5919 sea otter place
ste 200
carlsbad, CA 92010
MDR Report Key13805736
Report Number9616086-2022-00004
Device Sequence Number19841026
Product Code IMG
UDI-Device Identifier00888912296793
UDI-Public00888912296793
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K062354
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number2866
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-