| Model Number |
2866 |
| Health Effect - Clinical Code |
Superficial (First Degree) Burn (2685)
|
| Date of Event |
02/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was reported that the stim unit allegedly burned the patient.Djo is awaiting the return of this device.Additional reporting on this event will be provided as a supplemental report to this document if and when the device is returned for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the stim unit allegedly burned the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
After performing functional and electrical testing, the technician was unable to duplicate the reported complaint.The product met manufacturer specifications.Possible root causes include moisture or misapplication of the electrodes.
|
| |
|