H3, h6: the reported device was received for evaluation.An analysis of the customer provided images showed picture of the device box and a broken anchor on a inserter.A visual inspection revealed the device was returned with original packaging.The anchor is still on the inserter and the anchor¿s lower threads are broken off.There is only four upper threads left and the sutures have been cut from the lower portion of the device and are missing, the remainder of the sutures are still wrapped around the handle no other physical damage is visible to the device.A functional evaluation of the returned device found that mechanical portion of the device does work.Device passed all functional tests.It was determined the device contributed to the reported event.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors, which could have contributed to the complaint event, include an application of unintended or inappropriate or excessive force to the device.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found there are requirements for conformance and a certificate of material analysis is required.Based on the condition of the product material found during visual inspection it was determined the device damage was caused by the application of improper/excessive force.Additional material testing is not required.No containment or corrective actions are recommended at this time.One undated photo of the device was provided for review and confirms the reported breakage.Per case details, the broken anchor was retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.
|