• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER KIT 13F 20CM TL; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER KIT 13F 20CM TL; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED Back to Search Results
Model Number N/A
Medical Device Problem Codes Component Incompatible (1108); Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refx5423 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (refx5423) have been reported from the same facility.
 
Event or Problem Description
It was reported "the clinicians stated that they cannot feed the wire through the port upon insertion due to fraying of the wire itself." it was reported this occurred with two devices.This report addresses the first device.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), labeling, applicable manufacturing records, sample analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of unable to advance the guidewire through the catheter is confirmed and appears to be manufacturing related.One 20 cm trialysis catheter and one guidewire were returned for evaluation.Blood residue and evidence of use was noted on the catheter.The returned guidewire was inserted from the distal end of the power injectable lumen of the catheter.Resistance was noted 1 cm from the distal end of the catheter.The catheter was cut at the region resistance was experience in order to measure the lumen dimensional properties.The lumen size was found to be below manufacturing specification which is a likely factor to the difficult guidewire advancement; therefore, the complaint is confirmed.Bd is working closely with the supplier/manufacturing to prevent recurrence of the reported event.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported "the clinicians stated that they cannot feed the wire through the port upon insertion due to fraying of the wire itself." it was reported this occurred with two devices.This report addresses the first device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER KIT 13F 20CM TL
Common Device Name
CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13806598
Report Number3006260740-2022-00841
Device Sequence Number16661284
Product Code NIE
UDI-Device Identifier00801741041860
UDI-Public(01)00801741041860
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K133456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number5603200
Device Lot NumberREFX5423
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date05/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-