| Model Number |
1076538 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.During device evaluation, particle and water contamination was found.There was no report of serious patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Search Alerts/Recalls
|