| Model Number |
PM1110 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient presented to a scheduled procedure.During the procedure, the pacemaker's set screw would not tighten.The device was not used and another device was used.The patient condition was stable post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of set screw unable to fix the lead was not confirmed.Visual inspection of the set screw was normal with no anomalies.During analysis, a lead was inserted and removed multiple times with normal force, and the setscrew fixed the lead to the chamber without anomalies.Electrical, mechanical and longevity assessments were normal.
|
| |
|