ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804400-18
Medical Device Problem Codes Deflation Problem (1149); Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The stent remains in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported the procedure was to treat a heavily calcified, moderately tortuous lesion in the proximal left anterior descending artery.The 4.0x18 mm skypoint stent delivery system (sds) was not prepared outside of the body.The (sds) was advanced to the target lesion without issue.The sds was inflated once to nominal pressure.During deflation negative pressure was held for several seconds, several times, the balloon did not fully deflate.When attempting to remove the sds, the sds got stuck in the guide catheter since the balloon was withdrawn partially deflated.The sds and guide catheter were removed as a single unit.A new guide catheter and non-compliant balloon was used to complete the procedure.There were no adverse patient effects and no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.It was reported that the stent delivery system (sds) was prepped inside the anatomy.It should be noted that the xience skypoint, everolimus eluting coronary stent system (eecss), electronic instructions for use states: prepare an inflation device / syringe with diluted contrast medium.Attach an inflation device / syringe to the stopcock; attach it to the inflation port.With the tip down, orient the delivery system vertically.Open the stopcock to delivery system; pull negative for 30 seconds; release to neutral for contrast fill.Close the stopcock to the delivery system; purge the inflation device / syringe of all air.Repeat steps until all air is expelled.In this case, it is unknown if the ifu deviation contributed to the reported event.The investigation determined the reported difficult to remove appears to be related to the operational context of the procedure.Additionally, the investigation was unable to determine a conclusive cause for the reported deflation problem.There is no indication of a product quality issue with respect to manufacture, design, or labeling; therefore, no product-related corrective action will be implemented in this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13811479
Report Number2024168-2022-02838
Device Sequence Number14628203
Product Code NIQ
UDI-Device Identifier08717648233487
UDI-Public08717648233487
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2023
Device Model Number1804400-18
Device Catalogue Number1804400-18
Device Lot Number1082441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/28/2022
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-