• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 36MM +8; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 36MM +8; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71303608
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 10/12/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
*us legal* it was reported that, after a r3-tha in (b)(6) 2017, plaintiff presented metallosis.Plaintiff underwent revision surgery on (b)(6) 2021.Plaintiff outcome is unknown.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
*us legal* it was reported that, after a r3-tha in (b)(6) 2017, plaintiff presented metallosis.Plaintiff underwent revision surgery on (b)(6) 2021.Plaintiff outcome is unknown.
 
Event or Problem Description
It was reported that, after a primary right r3-tha surgery in (b)(6) 2017, the plaintiff presented metallosis, significant pain, edema, and mechanical complications.As a result, the plaintiff underwent revision surgery on (b)(6) 2021.During the revision, corrosion was found at the neck and head junction.The femoral component and the acetabular shell were well fixed and were left in place.The liner and the femoral head were explanted and replaced with a tha system from a competitor.The plaintiff tolerated the procedure well.
 
Additional Manufacturer Narrative
Internal reference number: (b)(4).
 
Additional Manufacturer Narrative
H10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that the reported pain, elevated cobalt and intraoperative findings of ¿slightly cloudy, yellow joint fluid, grossly thickened capsule, and corrosion at the head and neck junction may be consistent with the reported metallosis.However, the root cause of the reported clinical findings cannot be confirmed but based on the intraoperative findings of corrosion on the trunnion is a likely source of the metal debris.It cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that fracture of implant components has been identified as a possible adverse effect.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include damaged product, implant corrosion or wear.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HEAD 36MM +8
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13814683
Report Number1020279-2022-01229
Device Sequence Number2068652
Product Code JDG
UDI-Device Identifier03596010480002
UDI-Public03596010480002
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K022902
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71303608
Device Catalogue Number71303608
Device Lot Number17CM00646
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2022
Supplement Date Received by Manufacturer02/25/2022
03/21/2022
04/22/2022
Initial Report FDA Received Date03/18/2022
Supplement Report FDA Received Date03/21/2022
03/30/2022
04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71332752 / R3 0 DEG XLPE ACET LNR / 16LM14840.; 71335552 / R3 3 HOLE ACET SHELL 52MM / 17BM08020A.; 71356006 / ANTHOLOGY SO POROUS SIZE 6 / 16FM14196.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age77 YR
Patient SexFemale
-
-