Model Number 01.32.3239HCT
Health Effect - Clinical Code Pain (1994)
Date of Event 02/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 02 march 2022: lot 2001302: (b)(4) items manufactured and released on 12-may-2020.Expiration date: 2025-04-30.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Additional implant involved: ball heads: mectacer 01.29.204 biolox delta ceramic ball head 12/14 ø 32 size s -4 (k112115) lot.2001866.Batch review performed on 02 march 2022: lot 2001866: (b)(4) items manufactured and released on 07-july-2020.Expiration date: 2025-06-25.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.
 
Event or Problem Description
The patient came in for a post-op appointment reporting pain.The surgeon suspected that a loose cup was the cause of pain.At 1 year and 4 months after primary during the revision surgery, the surgeon found the cup was well fixed and decided to only revise the head and liner.The surgery was completed successfully.The cause of the pain is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER: MPACT FLAT PE HC LINER Ø32/C
Common Device Name
ACETABULAR PE HC LINER
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key13820177
Report Number3005180920-2022-00200
Device Sequence Number910933
Product Code LPH
UDI-Device Identifier07630030811845
UDI-Public07630030811845
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103721
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model Number01.32.3239HCT
Device Catalogue Number01.32.3239HCT
Device Lot Number2001302
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2022
Initial Report FDA Received Date03/19/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-