Model Number 60-5274-132
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
The customer reported that the device, 60-5274-132, stealth 32 lhook lap elec pkg, was being used during a total laparoscopic hysterectomy on (b)(6) 2022 when it was reported ¿that the customer found peeling of the l-hook electrode coating during surgery.During the operation, they found peeling of the l hook coating before using it for the patient, so they prepared an alternative device, but it was also confirmed to peel off.Therefore, surgery was performed with the third device and completed without any problem.¿.A second device, same lot, experienced the same type of malfunction; however, a third, same-like device was used to complete the procedure.There was no report of fragmentation contact to the patient.There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Event or Problem Description
The customer reported that the device, 60-5274-132, stealth 32 lhook lap elec pkg, was being used during a total laparoscopic hysterectomy on (b)(6) 2022 when it was reported ¿that the customer found peeling of the l-hook electrode coating during surgery.During the operation, they found peeling of the l hook coating before using it for the patient, so they prepared an alternative device, but it was also confirmed to peel off.Therefore, surgery was performed with the third device and completed without any problem.¿.A second device, same lot, experienced the same type of malfunction; however, a third, same-like device was used to complete the procedure.There was no report of fragmentation contact to the patient.There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.Update: (b)(6) 2022 - additional information was provided that peeling was found while using the device during surgery.
 
Additional Manufacturer Narrative
Received two 60-5274-132 in unoriginal packaging.Lot number was not verified.Performed a visual inspection, the complaint was confirmed.The black insulation is peeling off from the l-hook electrode.Based upon the evaluation and the photographs, the likely cause of this issue is the user activating the device continuously and at higher power than necessary.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 19 complaints, regarding 20 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised that these electrosurgical suction/irrigation instruments should only be used by surgeons trained in surgical endoscopy according to established hospital protocol.Misuse of this or any other electrosurgical device, can result in a patient injury.Read and become familiar with all instructions and directions before using this device.Always use the lowest possible settings to achieve the desired electrosurgical effect.Additionally, the ifu advises the user that this product is designed for non-continuous operation with a duty cycle of 10 seconds on and 30 seconds off.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTH 32 LHOOK LAP ELEC PKG
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
john berga
11311 concept blvd
largo, FL 33773
7273995358
MDR Report Key13823719
Report Number1320894-2022-00064
Device Sequence Number16664964
Product Code GEI
UDI-Device Identifier30653405010733
UDI-Public(01)30653405010733(17)261205(10)202112061
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K952204
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number60-5274-132
Device Catalogue Number60-5274-132
Device Lot Number202112061
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer04/22/2022
Initial Report FDA Received Date03/19/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-