C.R. BARD, INC. (BASD) -3006260740 DOT PROV 4F DL MAX 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by healthcare professional "measured the picc length before placing.Measured at 55cm- so no cutting necessary.Tried to thread the catheter-met resistance at 10 cm out and could not flush at that point.Pulled catheter out, at the sheath catheter twisted some.Tried to place again-same thing the catheter met resistance at 10 cm out and could not flush.Catheter removed to cut to mid line length to hold access.I pulled the wire back to trim catheter.Then noticed wire tip still in catheter.Pulled short piece of wire out of catheter.We had x-ray of arm/chest done-no fragments of wire seen." additional information received 02/24/2022: how was the break and/or retained segment discovered? piece of wire still in picc after removal.X-ray done to check for retained parts.Did the patient require surgery to remove the broken piece? no.Were all the missing/broken pieces retrieved and accounted for? yes.What type of imaging was used to verify broken piece in patient? x-ray done- no piece of wire seen.What type of catheter securement was utilized? no.At what point during device use (i.E.Insertion, infusion, removal) did the break occur? insertion.Any long-term patient harm? no.Was there any patient harm reported? no harm done.
 
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refs1619 showed one other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a broken stylet is confirmed; however, the exact cause is unknown.One 4 fr dual lumen powerpicc provena catheter with a t-lock and 3cg stylet inserted was returned for evaluation.An initial visual observation showed use residue on the returned sample.The returned stylet was observed to be broken into two pieces approximately 3.4 cm proximal to its distal tip.The epoxy tip of the stylet was observed present and intact on the distal segment and many bends were observed in the distal end of the stylet.A microscopic observation revealed the break sites in the stylet tubing were somewhat uneven with a rough surface texture and plastic deformation.Some tapering and delamination were observed in the stylet tubing at the break sites.The proximal break site was observed to be more tapered than the distal break site.The core wire of the stylet at the break site was observed to be tapered and curved slightly.The bends in the stylet tubing were observed to be between magnet interfaces.The observed fracture features were indicative of severe and/or repetitive stylet kinking, which likely occurred during the insertion process.However, the exact circumstances providing for that type of event were ultimately unknown.
 
Event or Problem Description
It was reported by healthcare professional "measured the picc length before placing.Measured at 55cm- so no cutting necessary.Tried to thread the catheter-met resistance at 10 cm out and could not flush at that point.Pulled catheter out, at the sheath catheter twisted some.Tried to place again-same thing the catheter met resistance at 10 cm out and could not flush.Catheter removed to cut to mid line length to hold access.I pulled the wire back to trim catheter.Then noticed wire tip still in catheter.Pulled short piece of wire out of catheter.We had x-ray of arm/chest done-no fragments of wire seen." additional information received 02/24/2022: how was the break and/or retained segment discovered? piece of wire still in picc after removal.X-ray done to check for retained parts did the patient require surgery to remove the broken piece? no.Were all the missing/broken pieces retrieved and accounted for? yes.What type of imaging was used to verify broken piece in patient? x-ray done- no piece of wire seen what type of catheter securement was utilized? no.At what point during device use (i.E.Insertion, infusion, removal) did the break occur? insertion any long-term patient harm? no.Was there any patient harm reported? no harm done.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PROV 4F DL MAX 3CG
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13825350
Report Number3006260740-2022-00874
Device Sequence Number16664974
Product Code LJS
UDI-Device Identifier00801741130366
UDI-Public(01)00801741130366
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS1274108D
Device Lot NumberREFS1619
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date03/19/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age75 YR
Patient SexMale
Patient Weight109 KG
-
-