| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported during the incoming inspection at (b)(6) warehouse in (b)(6).A team member found debris in the sterile package.No patient involvement.No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Foreign: country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned product identified that there is debris in the sterile packaging blister.The complaint has been confirmed by visual evaluation.Device history record (dhr) was reviewed and no discrepancies were found.The likely condition of the product when it left zimmer biomet control was non-conforming.The root cause of the reported event is the operator not following instructions during the manufacturing process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|