• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL MANUFACTURING .IMPRESS®.BRAIDED; INTERVENTIONAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL MANUFACTURING .IMPRESS®.BRAIDED; INTERVENTIONAL CATHETER Back to Search Results
Catalog Number 510038HH1
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device will not be returning for evaluation.A lot number was reported, and a review of the manufacturing history record is in progress.
 
Event or Problem Description
The customer reports that they had a tip detachment during stentgraft for an abdominal evar procedure.When an impress catheter was attempted to be withdrawn from a patient, the operator felt physical resistance, and upon removal, the tip of the catheter broke off, at the leg of the stentgraft, within abdominal aorta.The separation was at the border between the black and purple portions at the tip, and its cross section was found to be smooth.The detached tip was successfully retrieved by endovascular procedure with a snare catheter.No additional adverse event was reported.
 
Additional Manufacturer Narrative
The suspect device has been returned for evaluation.The device was examined visually and microscopically.Functional testing was performed.The complaint is confirmed.The root cause is attributed to the manufacturing process.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
.IMPRESS®.BRAIDED
Common Device Name
INTERVENTIONAL CATHETER
Manufacturer (Section D)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer Contact
bryson heaton bsn,rn.
1600 merit parkway
8012084662
MDR Report Key13825555
Report Number3010665422-2021-00125
Device Sequence Number1000019
Product Code DQO
UDI-Device Identifier00884450009116
UDI-Public884450009116
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2024
Device Catalogue Number510038HH1
Device Lot NumberE2065195
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer03/07/2022
Initial Report FDA Received Date03/19/2022
Supplement Report FDA Received Date03/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-