| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported ¿customer had a leaking picc.Leaking occurred today where the needle free connector meets the picc hub.Leakage is coming from the interface at clave (needleless connector) and picc hub on purple port.It was stated the clave ( needleless connector) was changed out to one from a different manufacturer and ended up with the same leakage.There is no indication of patient harm.".
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed and the cause was manufacturing-related.The product returned for evaluation was one 5fr d/l powerpicc catheter.Usage residues were observed throughout the sample.Needleless injection caps were attached to both luer adapters.An attempt to infuse water through the sample using a 12ml syringe revealed both lumens to be patent to infusion; however, during pressurization, leakage was observed at the junction between the purple luer adapter and the needleless injection cap.Inspection of the purple luer adapter using an iso taper gauge revealed that the gauge could not be fully inserted into the luer orifice.Microscopic inspection of the purple luer adapter revealed a protrusion in the luer material within the orifice.The mold defect appeared to prevent firm attachment of the injection cap, as well as full insertion of the taper gauge.The molding defect observed within the luer orifice prevented firm attachment with the needleless injection cap and resulted in leakage at the interface between the catheter and the cap.Photographs of the sample have been forwarded to the manufacturing site for further evaluation.A lot history review (lhr) of refx4493 showed seven other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported ¿customer had a leaking picc.Leaking occurred today where the needle free connector meets the picc hub.Leakage is coming from the interface at clave (needleless connector) and picc hub on purple port.It was stated the clave ( needleless connector) was changed out to one from a different manufacturer and ended up with the same leakage.There is no indication of patient harm.".
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed and the cause was manufacturing-related.The product returned for evaluation was one 5fr d/l powerpicc catheter.Usage residues were observed throughout the sample.Needleless injection caps were attached to both luer adapters.An attempt to infuse water through the sample using a 12ml syringe revealed both lumens to be patent to infusion; however, during pressurization, leakage was observed at the junction between the purple luer adapter and the needleless injection cap.Inspection of the purple luer adapter using an iso taper gauge revealed that the gauge could not be fully inserted into the luer orifice.Microscopic inspection of the purple luer adapter revealed a protrusion in the luer material within the orifice.The mold defect appeared to prevent firm attachment of the injection cap, as well as full insertion of the taper gauge.The molding defect observed within the luer orifice prevented firm attachment with the needleless injection cap and resulted in leakage at the interface between the catheter and the cap.Photographs of the sample have been forwarded to the manufacturing site for further evaluation.H3 other text : evaluation findings are in section h.11.
|
| |
|