| Model Number |
2941 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Pressure Sore/Ulcer (2326)
|
| Date of Event |
03/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient sustained a stage 4 pressure injury on an isoair support surface.Attempts are being made to gather more details from the user facility.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental is being filed to correct the model number and serial number of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient sustained a stage 4 pressure injury on an isoair support surface.Attempts are being made to gather more details from the user facility.
|
| |
|
Additional Manufacturer Narrative
|
|
A stryker staff design engineer evaluate the unit, where he found that the deflated central air bladders were due to kinking of the side bolster pressure sensor tube.The pump was returned to stryker and further evaluated, where it was determined that the pump properly prompted the disconnect air hose alarm.The patient sustained a stage 4 pressure injury on the surface, however, the patient was designated as high risk for a pressure injury.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient sustained a stage 4 pressure injury on an isoair support surface.
|
| |
|