|
|
| Model Number |
47-4828-016-02 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Hyperresponsive to stimuli (4538); Swelling/ Edema (4577)
|
| Date of Event |
02/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: part: 47235701704, lot: 63202115, distal lateral fibular plate right 4 holes 80 mm length.Part: 47234801635, lot: 63085306, cortical bone screw self-tapping hex head 3.5 mm diameter.16 mm length part: 47482801402, lot: 63202015, 2.7 mm locking screw 14 mm length.Part: 47234801835, lot: 63098016, cortical bone screw self-tapping hex head 3.5 mm diameter 18 mm length.Part: 47482801602, lot: 63457984, 2.7 mm locking screw 16 mm length.Part: 47482801602, lot: 63457984, 2.7 mm locking screw 16 mm length.Part: 47482801602, lot: 63488874, 2.7 mm locking screw 16 mm length.Part: 47235901638, lot: 63174173, 3.5 mm locking screw with 2.7 mm head 16 mm length.Report source: foreign- (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2022 - 00540, 0001822565 - 2022 - 00730, 001822565 - 2022 - 00731, 0001822565 - 2022 - 00732, 0001822565 - 2022 - 00734, 0001822565 - 2022 - 00735, 0001822565 - 2022 - 00736, 0001822565 - 2022 - 00784.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent right ankle open reduction internal fixation procedure.Subsequently swelling, pain, sensitivity, discoloration, and unable to wear shoes was reported with plate and all screws removed with no complications.No further information was provided.
|
| |
|
Event or Problem Description
|
|
No additional event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon further investigation, the device was found related to the reported event.Medwatch (0001822565-2022-00733-1) was submitted in error and the event remains reportable.Reported event was confirmed by review of medical records.Review of the available records identified the following: swelling, pain, sensitivity, discoloration, and unable to wear shoes.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
Upon receipt of additional information, it has been determined that the screws were not the cause of the corrosion, pain, or swelling and not device-related.This device did not cause or contribute to the reported event.The initial report was forwarded in error and should be voided.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt of additional information, it has been determined that the screws were not the cause of the corrosion, pain, or swelling and not device-related.This device did not cause or contribute to the reported event.The initial report was forwarded in error and should be voided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d2, d9, g3, g6, h1, h2, h11.The following section was corrected: d4(lot#).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned products identified discoloration on the fibular plate around three screw holes and one locking screw.The discolored screw was sent for further analysis.Eds semi-quantitative elemental analysis of the discolored regions on the screw head revealed foreign elements.It revealed c, n, o, na, mg, p, cl, k, and ca as the major foreign elements.The source of these elements could be from various potential contaminants including biological soft tissue and bone, dried biological fluids (i.E.Blood), bone cement (calcium phosphate-based), corrosion/metal oxide product or various decontamination solutions (i.E.Hospital detergents).Therefore, the source of these elements is inconclusive.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|
|
|