| Model Number |
11161AB-S |
| Medical Device Problem Code |
Intermittent Loss of Power (4016)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the customer had three incidents involving part # 11161ab-s (bugbee electrode).The item worked either intermittently or not at all, during our investigation of the event, the facility had no further information to provide.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.Per the customer the electrode was used with our 26006m but they also tried to troubleshoot with an aesculap monopolar cable.Our sales representative went to the account and tested the electrode with both cords and everything worked.
|
| |
|