| Model Number |
F2417H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an animal underwent an unknown procedure on (b)(6) 2022 and suture was used.During the procedure, the needle pulled off from the thread, without exerting excessive tension.Another like device was used to complete the procedure without any incident on the animal.There were no patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 4/25/2022.H3 investigational summary: the product was returned to ethicon inc for evaluation.Visual inspection was conducted on the returned sample.Visual analysis of the returned sample determined that one empty labeled winding former, a detached needle, and a suture piece of product code f2417h were received for evaluation.An incorrect swage was noted since the marks of the swage area were misaligned (one stake near the suture) and consequently, a needle pull of defect was caused.Based on the information currently available, incorrect swage was identified during the investigation of the sample received.This product issue will be addressed through ethicon inc¿s quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|