| Model Number |
989605410251 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Return of the suspect tilt swivel arm assembly is anticipated.Evaluation of the arm will be included in a follow-up report upon its return and investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported the monitor arm bracket broke off of their affiniti 50 ultrasound system while transporting the unit within the user facility.A philips service engineer ordered a replacement tilt swivel arm that was installed by the customer biomedical department to repair the system.No patient or user was harmed as a result of the issue.The system has been returned to service with no additional issues reported.
|
| |
|
Event or Problem Description
|
|
A customer reported the monitor arm bracket broke off of their affiniti 50 ultrasound system while transporting the unit within the user facility.A philips service engineer ordered a replacement tilt swivel arm that was installed by the customer biomedical department to repair the system.No patient or user was harmed as a result of the issue.The system has been returned to service with no additional issues reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Although the suspect arm was inadvertently discarded, a thorough investigation has been performed to identify the root cause of the reported issue, including a manufacturing evaluation and analysis of the involved monitor arm variant.The engineering team determined the failure resulted from an assembly error causing misaligned screw threads to shear away from a secured position.A design change that involves securing the monitor arm with a tab and slot instead sole reliance on screws has been implemented in january 2019 and prevents potential recurrence of this assembly error.
|
| |
|