Model Number Z339H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.The following information was requested, but unavailable: procedure name and date? no further information is available.Event date? no further information is available.What tissue was being approximated or sutured when it broke? no further information is available.What was used to complete the procedure? no further information is available.Events reported via: 2210968-2022-01923.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an known date and suture was used.During the procedure, suture breakage occurred 2 times in a row during use.Another like device was used to complete the surgery.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 4/4/2022.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational narrative: the product was returned to ethicon inc for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that a paper lid, an empty winning former, and a suture piece of product code z339h were received for evaluation, the suture piece was observed with marks on the surface and was cut appears to be by use of the surgical instrument.The product code z339h contains an absorbable suture and as the sample was received open, the time of exposure of the suture to the environment could not be determined and no functional test can be performed due to sample condition.The condition of the sample received indicates improper handling of the device.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational narrative: nineteen packets of product code z339h were returned to ethicon inc for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strands to detect any issue related to damaged or breakage suture and no defects were observed during the evaluation.A functional test was performed using an instron and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDSII VIO 27IN 2-0 S/A CT-1
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13827153
Report Number2210968-2022-01924
Device Sequence Number2083322
Product Code NEW
UDI-Device Identifier10705031060609
UDI-Public10705031060609
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Model NumberZ339H
Device Catalogue NumberZ339H
Device Lot NumberRGMACR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date03/19/2022
Supplement Report FDA Received Date04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-