| Model Number |
0234102102 |
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.All pieces were retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: instrument broken probable root cause: design: poor mechanical advantage design inadequate material selection for cutter (distal or proximal features) distal assembly of cutter susceptible to fragmentation break away pin too weak excessive interference between jaw and shaft (big dog only) process: inadequate assembly of cutter cutting edge inadvertently dulled during manufacturing use error the reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|