• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH LACTATE DEHYDROGENASE (LDH); NAD REDUCTION/NADH OXIDATION, LACTATE DEHYDROGENASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH LACTATE DEHYDROGENASE (LDH); NAD REDUCTION/NADH OXIDATION, LACTATE DEHYDROGENASE Back to Search Results
Model Number 2P56-22
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
The account generated (b)(6) architect ldh that repeated lower.A patient sample generated architect ldh of (b)(6) that repeated (b)(6).The account ldh normal expected range is 162 to 243 u/l.No specific patient information was provided.No impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The investigation included review of complaint activity, trending reports, device history records and labeling.Return testing was not completed as returns were not available.The customer retested the sample and generated the expected lower normal results.Additionally, the customer's quality controls were in range indicating the assay was performing as expected.Review of complaint and trending data did not identify any issues or trends.Device history records for the complaint lot were reviewed and did not identify any issues related to the complaint issue.Labeling was reviewed and sufficiently addresses the customer's issue.Based on this investigation, no systemic issue or deficiency was identified for the architect ldh reagent lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LACTATE DEHYDROGENASE (LDH)
Common Device Name
NAD REDUCTION/NADH OXIDATION, LACTATE DEHYDROGENASE
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key13827305
Report Number3002809144-2022-00099
Device Sequence Number2149946
Product Code CFJ
UDI-Device Identifier00380740161460
UDI-Public00380740161460
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/08/2022
Device Model Number2P56-22
Device Catalogue Number02P56-22
Device Lot Number14071UN21
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2022
Supplement Date Received by Manufacturer03/31/2022
Initial Report FDA Received Date03/19/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/07/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C8 PROC MOD, 01G06-11, C804340.; ARC C8 PROC MOD, 01G06-11, C804340.
Patient SexUnknown
-
-