| Model Number |
DBP-125MICRO145 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The diamondback 360 gen 2 peripheral orbital atherectomy system insructions for use manual states dissection is a potential adverse events that may occur and/or require intervention with use of the system.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|
| |
|
Event or Problem Description
|
|
Atherectomy was performed with a diamondback peripheral orbital atherectomy device (oad) in the peroneal artery via contralateral approach.Treatment was followed by imaging, which revealed a dissection at the bifurcation.Balloon angioplasty was applied for 15 minutes to resolve the dissection.The patient was in good condition and stable.
|
| |
|